Medical Surgical Bed recall
Medical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended to support and transport patients within the…
Why was Medical Surgical Bed recalled?
Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model 3005S3EX, Serial Number 120415087 - 120515468,
- Product
- Medical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended to support and transport patients within the…
- Recalling firm
- Stryker Medical Division of Stryker Corporation
- Quantity
- 1004
- Distribution
- Worldwide Distribution - Nationwide and Canada
- Recall date
- January 9, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)