FDAClass IIDecember 24, 2014

PKG, SPRING HANDLE, P/N 0250080237. Laparoscopic Manual recall

PKG, SPRING HANDLE, P/N 0250080237. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs…

Why was PKG, SPRING HANDLE, P/N 0250080237. Laparoscopic Manual recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

413144 414465 614433 716055 717057 717566 913220 913565 913566 913625 913763 914558 914559 914560 914654 914655 915026 915062 915269 915270 915430 915791 915895 1013200 1013201 1013277 1013393 1014635 1015163 1015164 1015774 1015863 1016125 1016405 1016407 1016555 1016556 1113333 1113334 1113335 1113340 1114205 1114544 1114545 1114546 1115529 1115530 1115650 1115652 1115653 1116111 1116113 1116460 1116461 1117176 1117177 1213128 1213258 1213259 1213973 1213975 1214001 1214735 1214736 1215298 1215299 1216023 1216024 1216228 1216229 1216288 1216721 1216722 1313330 1313331 1313332 1313805 1313806 1313947 1313948 1313972 1314203 1314204 1314472 1315785 1315786 1315787 1410413

Product
PKG, SPRING HANDLE, P/N 0250080237. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)