FDAClass IIJanuary 20, 2016

UniAxial Pedicle Screw & UniAxial Reduction Pedicle recall

UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the…

Why was UniAxial Pedicle Screw & UniAxial Reduction Pedicle recalled?

Response Spine UniAxial Pedicle & UniAxial Reduction Pedicle Screws may have been manufactured incorrectly, thus preventing the driver from fully engaging into the hexagonal drive feature of the screw.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Item number; lot; product description 00-1300-1735; 44003; 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 35MM. 00-1300-0735; 44601; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 35MM. 00-1300-0740; 44602; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0740; 43926; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0740; 44252; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0745; 43927; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 45MM. 00-1300-0750; 44619; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 50MM. 00-1300-0620; 44623; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 20MM. 00-1300-0640; 44617; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM. 00-1300-0640; 44884; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM. 00-1300-1745; 44005; 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 45MM. 00-1300-0635; 44677; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 35MM.

Product
UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the…
Recalling firm
OrthoPediatrics Corp
Quantity
102
Distribution
US Distribution to states of: AL, AZ, CA, DE, FL, GA, HI, KY, LA, MI, MO, MN, MS, NC, NJ, NY, OH, OR, TN, and TX.
Recall date
January 20, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)