FDAClass IIDecember 11, 2024

EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 recall

EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190

Why was EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 recalled?

High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Number: GIF-1TH190; UDI: 04953170343360, 04953170437014; Serial Number Range: Before 2446103 or After 2500453

Product
EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
Recalling firm
Olympus Corporation of the Americas
Quantity
1,770 units
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
Recall date
December 11, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)