PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O recall
PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O TROCAR, P/N 0250080171. Laparoscopic Manual Instruments are intended for cutting, holding, grasping…
Why was PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O recalled?
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
35499 38198 38758 39650 39650N 40110 40250 43627 44955 44976 45866 46024 46320 46396 46710 46802 47210 47389 47513 47750 47970 48066 48398 48762 48965 49494 49498 49612 49834 50168 50430 50722 50855 50920 51536 52647 52919 53043 53136 53515 53795 53972 55872 58336 58821 59615 60227 60752 62482 64133 64678 64965 65640 67422 67737 68342 69272 71077 75257 75918 76080 76460 76831 76960 77711 78491 79425 82007 82916 83532 83926 85069 85467 85909 86149 86265 86430 86539 86869 87496 89088 89441
- Product
- PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O TROCAR, P/N 0250080171. Laparoscopic Manual Instruments are intended for cutting, holding, grasping…
- Recalling firm
- Stryker Endoscopy
- Quantity
- 180,573 units total
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Recall date
- December 24, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)