FDAClass IIJanuary 13, 2016

Oncentra External Beam - VMAT - Radiation therapy recall

Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software…

Why was Oncentra External Beam - VMAT - Radiation therapy recalled?

When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Versions 3.3 and higher

Product
Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software…
Recalling firm
Elekta, Inc.
Quantity
154
Distribution
Worldwide Distribution - US Nationwide in the states of CA, NY, PA, and the countries of Austria, Belarus, Canada, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Netherlands, Norway, Poland, South Korea, Spain, Sweden and United Kingdom.
Recall date
January 13, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)