Asante Conset Infusion Set with 110cm tubing & 6mm recall
Asante Conset Infusion Set with 110cm tubing & 6mm cannula packaged in 5-packs, singles,or inside a Combination Kit (FG-5246-4) Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.
Why was Asante Conset Infusion Set with 110cm tubing & 6mm recalled?
Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Asante Catalog number:FG-4246-5, FG-4246T; Unomedical Item number:43-110-5266; Lot numbers: 0230128 5005803 5030168.
- Product
- Asante Conset Infusion Set with 110cm tubing & 6mm cannula packaged in 5-packs, singles,or inside a Combination Kit (FG-5246-4) Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.
- Recalling firm
- Asante Solutions, Inc.
- Quantity
- 454 kits
- Distribution
- Nationwide Distribution.
- Recall date
- January 8, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)