FDAClass IIDecember 11, 2024

Atellica¿ CH ¿2-Microglobulin recall

Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma. Siemens…

Why was Atellica¿ CH ¿2-Microglobulin recalled?

Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 00630414595566 All in-date and future Atellica CH B2M reagent lots

Product
Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma. Siemens…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
11330 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hungary, India, Iran, Israel, Italy, Ivory Coast, Japan, Jordan, Kuwait, Mexico, Morocco, Oman, P.R. China, Pakistan, Paraguay, Poland, Portugal, Saudi Arabia, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Switzerland, Thailand, Turkey, U.A.E., Uganda, United Kingdom, Uruguay, Uzbekistan, Vietnam, Zaire.
Recall date
December 11, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls