FDAClass IIDecember 24, 2014

PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080121 recall

PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080121 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…

Why was PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080121 recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial/Lot 3282011 06CON06112 817085 0817438/0817442 0817439/0817442 0913442/0913326 0913442/0913523 0913544/0913523 0913581/0913523 0913812/0817492 0913812/0914227 0913812/0915061 0914894/1015779 0915431/1016063 1016560/1016063 1016560/1016826 1016847/1117011 1115970/1117011 1115971/1117011 1116463/1216149 1216408/1216149 1216409/1216149 1216409/1313046 1216409/1313047 1216409/1313048 1216409/1313946 1216720/1313946 913442/913523

Product
PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080121 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls