Maquet HCU 30 Coolant Type R134A Filling weight recall
Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate…
Why was Maquet HCU 30 Coolant Type R134A Filling weight recalled?
Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
HCU 30 Base Unit 100 - 120 V, Model Number 704629/70103.4371 HCU 30 Base Unit 200 - 240 V, Model Number 704631/70103.4642 HCU 30 Base Unit Canadian/US, Model Number 705501/70103.4653 Heater Cooler Unit 30 with Serial Number <93101480 The affected power supply boards have serial numbers less than 107006001.
- Product
- Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate…
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Quantity
- 100 Units (distributed in United States)
- Distribution
- United States Nationwide Distribution
- Recall date
- January 13, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)