ADULT Radiotransparent Electrode recall
ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart…
Why was ADULT Radiotransparent Electrode recalled?
Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All Lots. Catalog / Part Number C100-PHILIPS
- Product
- ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart…
- Recalling firm
- Heart Sync, Inc.
- Quantity
- sold in boxes of ten electrodes: 4663.8 boxes; 46638 units total
- Distribution
- Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea.
- Recall date
- December 31, 2014
- Classification
- Class I
- Status
- Terminated
- Agency
- FDA (device)