FDAClass IIJanuary 13, 2016

Remote Patient Monitoring System recall

Remote Patient Monitoring System, MDP 2000 Series, Part Numbers: MDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is intended for use in conjunction…

Why was Remote Patient Monitoring System recalled?

Testing at customer site showed unit Remote Patient Monitoring System MDP2040-0100 in a continuous trap condition, not allowing system to reset and reboot. Two customers recently complained of the system freezing and it could only be restarted if the user re-applied power.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial numbers: 6533-6537, 6561, 6566-6581, 6539, 6540, 6558-6560, 6562, 6508-6528, 6563-6565, 6529, 6531, 6532, 6541, 6551, 6556, and 6557.

Product
Remote Patient Monitoring System, MDP 2000 Series, Part Numbers: MDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is intended for use in conjunction…
Recalling firm
Vidco, Inc.
Quantity
59 units
Distribution
US Nationwide distribution in the states of AZ, CA, MD, NM, NJ, and OH.
Recall date
January 13, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls