FDAClass IIDecember 4, 2019

Guidezilla II Guide Extension Catheter Sterilized using recall

Guidezilla II Guide Extension Catheter Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493933515060, 6F; 2. REF/Catalog No H7493933515070, 7F.

Why was Guidezilla II Guide Extension Catheter Sterilized using recalled?

Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot / GTIN Number / Use By: 1. 23485544 / 8714729939450 / 2021-02-10, 23933296 / 8714729939450 /2021-05-10, 23947883 / 8714729939450 / 2021-05-12, 24137846 / 8714729939450 / 2021-06-17. 2. Lots: 21305995 / 8714729939474 / 2019-09-25, 24201181 / 8714729939474 / 2021-06-30.

Product
Guidezilla II Guide Extension Catheter Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493933515060, 6F; 2. REF/Catalog No H7493933515070, 7F.
Recalling firm
Boston Scientific Corporation
Quantity
14,145 devices total
Distribution
Nationwide including Washington D.C. and Puerto Rico. OUS to include Canada.
Recall date
December 4, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls