FDAClass IINovember 26, 2025

Brand Name: Sterilmed Reprocessed Imaging Catheter recall: Sterility Concern

Brand Name: Sterilmed Reprocessed Imaging Catheter Product Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR USE WITH SIEMENS SYSTEM…

Why was Brand Name: Sterilmed Reprocessed Imaging Catheter recalled?

Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Code: GTIN: 10888551004732 LOT# SERIAL#: 2188147 131065 2207154 916179 2220232 1050886 2234795 1170283

Product
Brand Name: Sterilmed Reprocessed Imaging Catheter Product Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR USE WITH SIEMENS SYSTEM…
Recalling firm
Sterilmed, Inc.
Quantity
4
Distribution
United States Only: Kentucky, Maryland, Massachusetts, Mississippi, New Jersey, Pennsylvania, Texas, Utah
Recall date
November 26, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls