Zimmer Title 2 Bone Screw Remover Instrument recall
Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius TITLE 2…
Why was Zimmer Title 2 Bone Screw Remover Instrument recalled?
Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery up to 30 minutes.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot: L75066
- Product
- Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius TITLE 2…
- Recalling firm
- Zimmer, Inc.
- Quantity
- 14 devices
- Distribution
- US nationwide distribution.
- Recall date
- December 17, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)