DABRA Laser recall
DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing…
Why was DABRA Laser recalled?
Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Number RA-308 Product/Catalog Number: 1200-0001-01 Serial Numbers: RA00027, RA00046, RA00049, RA00050, RA00054, RA00056, RA00057, RA00058, RA00064, RA00069, RA00073, RA00081, RA00083, RA00084, RA00085, RA00089, RA00097, RA00098, RA00114, RA00115, RA00120, RA00121, RA00125, RA00127, RA00135, RA00138, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177.
- Product
- DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing…
- Recalling firm
- Ra Medical Systems, Inc.
- Quantity
- 36 devices
- Distribution
- US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.
- Recall date
- December 23, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)