FDAClass IIDecember 23, 2020

Microsurgical Operating Room Tables recall

Microsurgical Operating Room Tables;7300BX (18-351-70).

Why was Microsurgical Operating Room Tables recalled?

Unintended tilting of the Tabletop MST 7300BX and incorrect operation due to infrared operating signal errors of Tabletops MST 7300B and MST 7300BX.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Tilt Movement-Leaking Valve (7300BX-18-351-70) serial numbers: MUC-20163-0001, MUC-2016Z-0002, MUC-2016Z-0003, MUC-20177-0005, MUC-20177-0007, MUC-20177-0008, MUC-20177-0009, MUC-20189-0017 & MUC-2018Y-0018. Unintended movement due to infrared signal errors (7300BX-18-351-70) serial numbers: MUC-20163-0001, MUC-2016Z-0002, MUC-2016Z-0003, MUC-20177-0005, MUC-20177-0007, MUC-20177-0008, MUC-20177-0009, MUC-20189-0017, MUC-2018Y-0018 & MUC-2018Y-0019.

Product
Microsurgical Operating Room Tables;7300BX (18-351-70).
Recalling firm
Mizuho America, Inc.
Quantity
10
Distribution
US: AZ, CA, FL, MO, NC, OH, PA, TX, and WI. OUS: Greece, Israel, Japan, and Ukraine.
Recall date
December 23, 2020
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls