Case Carton: BD Oral Dispensing Syringe recall
Case Carton: BD Oral Dispensing Syringe 1 mL Clear with Tip Cap BD, Franklin Lakes, NJ 07417 USA Made in USA www.bd.com BD, Laagstraat 57, B-9140…
Why was Case Carton: BD Oral Dispensing Syringe recalled?
BD received a complaint report for mixed syringe tips (oral and non-oral syringe tips) within the same lot.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Device Listing Number - R060456 REF Lot # 305217 2076404 305217 2055156 305217 2020057 305217 2020056 305219 1346419
- Product
- Case Carton: BD Oral Dispensing Syringe 1 mL Clear with Tip Cap BD, Franklin Lakes, NJ 07417 USA Made in USA www.bd.com BD, Laagstraat 57, B-9140…
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 2,786,000 units
- Distribution
- (USA) nationwide
- Recall date
- December 26, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)