STEPHANIX D2RS Digital Dynamic Remote recall
STEPHANIX D2RS Digital Dynamic Remote Systemhttp://cts.fda.gov/division-tracking/images/trash.png Product The D2RS Digital Dynamic Remote System is…
Why was STEPHANIX D2RS Digital Dynamic Remote recalled?
The firm has detected a potential risk using the command. After releasing the command, the movement of the table may continue instead of stopping.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
DHJ008, DIG010, DIH012, DIH014, DIK022, DIK024, DIL025, DIL030, DJK051, DJK052, DKI032, DKI033, DKJ047, DKK051, DKK052, DLA001, DML062
- Product
- STEPHANIX D2RS Digital Dynamic Remote Systemhttp://cts.fda.gov/division-tracking/images/trash.png Product The D2RS Digital Dynamic Remote System is…
- Recalling firm
- Stephanix
- Quantity
- 17 units
- Distribution
- US Nationwide Distribution in the states of FL, IN, WV, TN, MO, WI, CA, FL, GA, PA
- Recall date
- December 12, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)