TD60 Transmitter battery charger used with the recall
TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
Why was TD60 Transmitter battery charger used with the recalled?
A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All battery chargers with part number 115-030108-00
- Product
- TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
- Recalling firm
- Mindray DS USA, Inc. dba Mindray North America
- Quantity
- 203
- Distribution
- Distributed in 26 states: AL, CA, CO, FL, GA, IA, IN, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, WA, and WY; and Puerto Rico.
- Recall date
- February 14, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)