Autosampler recall
Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
Why was Autosampler recalled?
There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model/UDI-DI/Serial Number: SIL-20ACXR CL/04540217420131/L20995250001, L20995250002, L20995250003, L20995250004, L20995250005, L20995250006, L20995250007, L20995250008, L20995350010, L20995350011, L20995350012, L20995350013, L20995350014, L20995350015, L20995450016, L20995450017, L20995450018, L20995450019, L20995450020, L20995450021, L20995450022, L20995450023, L20995450024, L20995450025, L20995450026, L20995550027, L20995550028, L20995550029, L20995550030, L20995650031, L20995650032, L20995650033, L20995650034, L20995650035, L20995650036, L20995650037, L20995650038, L20995650039, L20995650040, L20995650041, L20995650042, L20995650043, L20995650044, L20995650045, L20995650046, L20995560001, L20995660002, L20995660003, L20995660004, L20995660005, L20995660006, L20995660007, L20995660008, L20995660009, L20995660010, L20995660011, L20995660012, L20995660013, L20995660014, L20995660015, L20995660016, L20995660017, L20995660018, L20995660019, L20995660020, L20995660021, L20995660022, L20995660023, L20995660024, L20995660025, L20995660026, L20995660027, L20995660028, L20995660029, L20995660030, L20995660031, L20995660032, L20995660033, L20995660034, L20995660035, L20995660036, L20995660037, L20995660038, L20995660039, L20995660040, L20995660041, L20995660042, L20995660043, L20995660044, L20995660045, L20995660046, L20995660047, L20995660048, L20995660049, L20995660050, L20995660051, L20995660052, L20995660053, L20995660054, L20995660055, L20995660056, L20995660057, L20995250009, L20995270001, L20995270002, L20995270003, L20995270004, L20995270005, L20995370006, L20995370007, L20995370008, L20995370009, L20995370010, L20995370011, L20995370012, L20995370013, L20995370014, L20995370015, L20995370016, L20995370017, L20995370018, L20995370019, L20995370020, L20995370021, L20995370022, L20995370023, L20995370024, L20995370025, L20995470028, L20995770032. SIL-20ACHT CL/04540217420124/L20985350001, L20985450002, L20985750003, L20985750004. SIL-30AC CL/04540217420155/L21005250001, L21005250002, L21005250003, L21005250004, L21005250005, L21005250006, L21005250007, L21005350008, L21005350009, L21005350010, L21005350011, L21005350012, L21005350013, L21005350014, L21005350015, L21005350016, L21005350017, L21005350018, L21005350019, L21005350020, L21005350021, L21005350022, L21005350023, L21005350024, L21005450025, L21005550026, L21005550027, L21005650028, L21005650029, L21005750030, L21005750031, L21005570001, L21005570002, L21005670003, L21005670004, L21005670005, L21005670006, L21005670007, L21005670008, L21005770009, L21005770010, L21005770011, L21005770012, L21005770013, L21005770014.
- Product
- Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
- Recalling firm
- Shimadzu Corp. Analytical & Measuring Instruments Div.
- Quantity
- 179
- Distribution
- Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.
- Recall date
- December 4, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)