Zimmer Natural Nail System Tear Drop Guide Wire 2.4 mm recall
Zimmer Natural Nail System Tear Drop Guide Wire 2.4 mm Diameter, 70 cm length, Item Number 47-2490-098-01 Product Usage: The Ball Tip and Tear Drop…
Why was Zimmer Natural Nail System Tear Drop Guide Wire 2.4 mm recalled?
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All Lot Numbers Expiry Date Before March 2022
- Product
- Zimmer Natural Nail System Tear Drop Guide Wire 2.4 mm Diameter, 70 cm length, Item Number 47-2490-098-01 Product Usage: The Ball Tip and Tear Drop…
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 1,631 units
- Distribution
- Nationwide within U.S.A and Puerto Rico; International: Australia, Brazil, Canada, China, Germany, India, Japan, Netherlands & Singapore
- Recall date
- February 14, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)