FDAClass IIFebruary 9, 2022

Vagus Nerve Stimulation recall

Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and Model…

Why was Vagus Nerve Stimulation recalled?

A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: 05425025750115. Serial Numbers: 119441, 119442, 119443, 119452, 119453, 119454, 119455, 119456, 119457, 119458

Product
Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and Model…
Recalling firm
LivaNova USA, Inc.
Quantity
10
Distribution
US: NE, MO, AR, WV, KS, MA, IN
Recall date
February 9, 2022
Classification
Class II
Status
Terminated
Agency
FDA (device)