FDAClass IIDecember 25, 2013

Broncho Flexible Fiberscope recall

Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree…

Why was Broncho Flexible Fiberscope recalled?

KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

C1140, C1140E, C1160, C1160E, QKC1750, QKC1750E, QKC1694, QKC1694E

Product
Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree…
Recalling firm
Karl Storz Endoscopy America Inc
Quantity
141 units
Distribution
Nationwide Distribution.
Recall date
December 25, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)