Varian On-Board Imager Advanced Imaging System. The recall
Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to…
Why was Varian On-Board Imager Advanced Imaging System. The recalled?
Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Part Number P1003290001
- Product
- Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to…
- Recalling firm
- Varian Medical Systems, Inc.
- Quantity
- 32 units
- Distribution
- Worldwide Distribution.
- Recall date
- December 24, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)