Mammomat Inspiration system: Product Usage: mammography recall
Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical…
Why was Mammomat Inspiration system: Product Usage: mammography recalled?
It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
model number: 10140000
- Product
- Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical…
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 3
- Distribution
- US Nationwide Distribution in the state of: WV, NY, and FL..
- Recall date
- December 17, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)