FDAClass IIDecember 20, 2023

Iconix Knotless 1.4mm Anchor with XBraid S recall

Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The…

Why was Iconix Knotless 1.4mm Anchor with XBraid S recalled?

Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Product Code: 3911-7 14-520 UDI-DI Code: 00840277402172 Lot Number: 23080433 Product Code: 3911-7 14-521 UDI-DI Code: 00840277402189 Lot Number: 23081617

Product
Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The…
Recalling firm
Riverpoint Medical, LLC
Quantity
999 devices
Distribution
U.S. Nationwide distribution in the state of CA.
Recall date
December 20, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)