Micro Screen 6 Panel Cup with Adult recall
Micro Screen 6 Panel Cup with Adult (AU), Item No. MSCA-6MBAU
Why was Micro Screen 6 Panel Cup with Adult recalled?
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
G140256 G140281 G140307 G140340 G140418 G140419 G150135 G150141 G150187 G150257
- Product
- Micro Screen 6 Panel Cup with Adult (AU), Item No. MSCA-6MBAU
- Recalling firm
- Ameditech Inc
- Quantity
- 4,646 kits
- Distribution
- Nationwide Distribution.
- Recall date
- December 30, 2015
- Classification
- Class III
- Status
- Terminated
- Agency
- FDA (device)