FDAClass IIJanuary 26, 2022

Medichoice Infant Heel Warmer recall: Labeling Error

Medichoice Infant Heel Warmer, Reference # 989805643721 1079906

Why was Medichoice Infant Heel Warmer recalled?

After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All units are impacted

Product
Medichoice Infant Heel Warmer, Reference # 989805643721 1079906
Recalling firm
Philips North America Llc
Quantity
2,925,800 units
Distribution
US Nationwide and Canada.
Recall date
January 26, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls