Epix Electrosurgical Probes with Smoke Evacuation recall
Epix Electrosurgical Probes with Smoke Evacuation, Angled L-Hook Tip, 5mm x 42 cm REF CW002 QTY: 1 - Product Usage: The Epix Electrosurgical Probe…
Why was Epix Electrosurgical Probes with Smoke Evacuation recalled?
Unintentional shifting of the tip insulation of the firm's electrosurgical probes.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Number: CW002 Lot Numbers: 1336418 and 1340803 Model # Lot # UDI Code CW002 1336418 (01)00607915136147(17)210912(30)01(10)1336418 CW002 1340206 (01)00607915136147(17)211029(30)01(10)1340206 TA211 1340118 (01)20607915139050(17)210812(30)01(10)1340118 TA211 1342169 (01)20607915139050(17)210812(30)01(10)1342169
- Product
- Epix Electrosurgical Probes with Smoke Evacuation, Angled L-Hook Tip, 5mm x 42 cm REF CW002 QTY: 1 - Product Usage: The Epix Electrosurgical Probe…
- Recalling firm
- Applied Medical Resources Corp
- Quantity
- 28 units Domestically - 112 units Internationally
- Distribution
- Worldwide distribution - US Nationwide distribution and countries of Belgium; Germany; Switzerland; Spain; Finland; France; Great Britain; Israel; India; Italy; Lebanon; Netherlands; and Portugal.
- Recall date
- November 27, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)