Defibtech DDP-2002: Pediatric Defibrillation Pads For recall
Defibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, Lifeline ECG, Lifeline View (DDU-2XXX…
Why was Defibtech DDP-2002: Pediatric Defibrillation Pads For recalled?
Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life.there is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI: 00815098020454, DDP-2002, pediatric defibrillation pads, single pair; 10815098020451, DDP-2002, pediatric defibrillation pads, single pair configuration; 40815098020452, DDP-2002, pediatric defibrillation pads, 20 pairs
- Product
- Defibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, Lifeline ECG, Lifeline View (DDU-2XXX…
- Recalling firm
- Defibtech, LLC
- Quantity
- 22,294 total
- Distribution
- Ireland
- Recall date
- December 4, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)