Siemens Atellica CH Reaction Ring Cuvette Segment-has recall
Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and present them to each…
Why was Siemens Atellica CH Reaction Ring Cuvette Segment-has recalled?
A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or elevated QC and patient results for seven (7) analytes: Ammonia, Acetaminophen, Alanine Aminotransferase (ALT), Hemoglobin A1c, Creatinine Kinase (CK), Cystatin C, and Salicylate.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot # UDI N1527024 (01)00630414596143(10)N1527024(17)20291002; N1526424 (01)00630414596143(10)N1526424(17)20291001; N1525924 (01)00630414596143(10)N1525924(17)20290924 ; N1525324 (01)00630414596143(10)N1525324(17)20290918 ; N1524224 (01)00630414596143(10)N1524224(17)20290911; N1523724 (01)00630414596143(10)N1523724(17)20290904; N1523024 (01)00630414596143(10)N1523024(17)20290827; N1521824 (01)00630414596143(10)N1521824(17)20290814; N1521424 (01)00630414596143(10)N1521424(17)20290808; N1519324 (01)00630414596143(10)N1519324(17)20290719; N1518924 (01)00630414596143(10)N1518924(17)20290716; N1518224 (01)00630414596143(10)N1518224(17)20290710; N1518324 (01)00630414596143(10)N1518324(17)20290710; N1516224 (01)00630414596143(10)N1516224(17)20290621; N1515624 (01)00630414596143(10)N1515624(17)20290612; N1514924 (01)00630414596143(10)N1514924(17)20290605; N1514124 (01)00630414596143(10)N1514124(17)20290530; N1513324 (01)00630414596143(10)N1513324(17)20290522; N1512824 (01)00630414596143(10)N1512824(17)20290515; N1512424 (01)00630414596143(10)N1512424(17)20290510; N1511224 (01)00630414596143(10)N1511224(17)20290506; N1510724 (01)00630414596143(10)N1510724(17)20290423; N1510124 (01)00630414596143(10)N1510124(17)20290419.
- Product
- Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and present them to each…
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 9729 units
- Distribution
- US Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Congo, Republic C¿te dIvoire Croatia Cyprus Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Iran, Islamic Republic of Iraq Ireland Israel Italy Japan Jordan Kenya Kuwait Latvia Libya Malaysia Mexico Moldova, Rep. of Morocco Nepal Netherlands Norway Oman Pakistan Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vatikancity Vietnam
- Recall date
- November 27, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)