FDAClass IIJanuary 19, 2022

Flat Panel Detector recall

Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F, and INFX-8000H.

Why was Flat Panel Detector recalled?

Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

SID: 30039059, Alphenix System Serial Number: BGC20Y2010, & Flat Panel Detector Serial Number: C1B20Y2215 SID: 30030612, Alphenix System Serial Number: B8D20Z2001, & Flat Panel Detector Serial Number: C1B20Z2218; SID: 30030612, Alphenix System Serial Number: B8D20Z2001, & Flat Panel Detector Serial Number: C1B20Z2219; SID: 30046247, Alphenix System Serial Number: B1B20Z2001, & Flat Panel Detector Serial Number: C1B20Z2221; SID: 30064041, Alphenix System Serial Number: BGD2172011, & Flat Panel Detector Serial Number: C1B2172234; SID: 30065145, Alphenix System Serial Number: B1E2172002, & Flat Panel Detector Serial Number: C1B2172235; SID: 30058149, Alphenix System Serial Number: B1E2152001, & Flat Panel Detector Serial Number: C1B2152230; SID: 30002089, Alphenix System Serial Number: BHC20Z2003, & Flat Panel Detector Serial Number: C1B20Z2217.

Product
Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F, and INFX-8000H.
Recalling firm
Canon Medical System, USA, INC.
Quantity
8
Distribution
US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.
Recall date
January 19, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)