AutoMate 1250 recall: Contamination
AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with Aliquot Module - UDI…
Why was AutoMate 1250 recalled?
Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) .
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All units containing software Version(s) 4.1.1 and up
- Product
- AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with Aliquot Module - UDI…
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 83 automate systems
- Distribution
- US: NY, VA OUS: Austria Belgium Czech Republic France Germany Israel Italy Lithuania Mayotte Romania Slovakia South Africa Spain Sweden Switzerland United Kingdom
- Recall date
- November 27, 2019
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)