Covidien Signia" Small Diameter Curved Tip Intelligent recall
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;
Why was Covidien Signia" Small Diameter Curved Tip Intelligent recalled?
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
CFN: BOX06170V4; UDI-DI: 00763000988975; Lot numbers: 0231961695;
- Product
- Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;
- Recalling firm
- Covidien, LP
- Quantity
- 2 units (OUS only)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
- Recall date
- November 19, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)