Plum A+ Single Channel Infusion Pumps with Hospira recall
Plum A+ Single Channel Infusion Pumps with Hospira Mednet Software and a Connectivity Engine peripheral module that provides wired Ethernet and…
Why was Plum A+ Single Channel Infusion Pumps with Hospira recalled?
The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
pump list numbers 20679 and 20792, and module list number 20677; all serial numbers
- Product
- Plum A+ Single Channel Infusion Pumps with Hospira Mednet Software and a Connectivity Engine peripheral module that provides wired Ethernet and…
- Recalling firm
- Hospira Inc.
- Quantity
- N/A
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico, and the countries of Australia, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Lavenia, South Africa, Spain, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates and Vietnam.
- Recall date
- December 19, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)