Mako Integrated Cutting System recall: Software Issue
Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product…
Why was Mako Integrated Cutting System recalled?
Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
(01) 00848486030193 (21)<serial number>, Serial numbers: 4200052, 4200095, 4200112, 4200117, 4200119, 4200167, 4200190, 4200193, 4200202, 4200203, 4200204, 4200226, 4200233, 4200235, 4200248, 4200250, 4200257, 4200269, 4200270, 4200288, 4200293, 4200305, 4200350, 4200373, 4200384, 4200387, 4200391, 4200397, 4200398, 4200400, 4200409, 4200436, 4200457, 4200459, 4200477, 4200488, 4200512, 4200540, 4200548, 4200550, 4200558, 4200624, 4200645, 4200653, 4200666, 4200717, 4200766, 4200774, 4200785, 4200789, 4200801, 4200806, 4200807, 4200816, 4200843, 4200856, 4200868, 4200904, 4200917, 4201000, 4201040, 4201068, 4201089, 4201102, 4201103, 4201121, 4201167, 4201218, 4201236, 4201238, 4201253, 4201296, 4201302, 4201366, 4201367, 4201376, 4201385, 4201408, 4201412, 4201442, 4201452, 4201454, 4201477, 4201490, 4201630, 4201661, 4201714, 4201752, 4201829, 4201831, 4201851, 4202009, 4202020, 4202038, 4202044, 4202062, 4202063, 4202122, 4202146, 4202162, 4202175, 4202315, 4202333, 4202626, 4202631, 4202659, 4202729, 4202742, 4202882, 4202987, 4203180, 4203198, 4203289, 4203424, 4203441, 4203649, 4203709, 4203728, 4203730, 4203731, 4203736, 4203807, 4203826, 4204015, 4204164, 4204389, 4204448, 4204539, 4204553, 4204577, 4204673, 4204812, 4204934, 4205093, 4205190, 4205380, 4205426, 4206219, 4206271, 4209023, 4209025, 4209044, 4209045, 4209046, 4209065, 4209066, 4209067, 4209068, 4209071, 4209149, 4209151, 4209155, 4209156, 4209157, 4209162, 4209164, 4209166, 4209167, 4209171, 4209185, 4209189, 4209210, 4209211, 4209213, 4209220, 4209244, 4209250, 4209252, 4209255, 4209256, 4209259, 4209261, 4209262, 4209267, 4209268, 4209271, 4209298, 4209299, 4209303, 4209304, 4209306, 4209308, 4209312, 4209317, 4209318, 4209345, 4209346, 4209357, 4209360, 4209363, 4209364, 4209365, 4209387, 4209388, 4209389, 4209398, 4209441, 4209443, 4209460, 4209463, 4209464, 4209466, 4209471, 4209472, 4209473, 4209474, 4209475, 4209476, 4209477, 4209486, 4209487, 4209491, 4209493, 4209494, 4209495, 4209496, 4209500, 4209519, 4209522, 4209526, 4209529, 4209531, 4209532, 4209533, 4209535, 4209541, 4209542, 4209543, 4209546, 4209597, 4209598, 4209599, 4209600, 4209601, 4209602, 4209604, 4209605, 4209606, 4209607, 4209608, 4209609, 4209610, 4209611, 4209612, 4209613, 4209614, 4209615, 4209616, 4209621, 4209628, 4209632, 4209634, 4209635, 4209637, 4209639, 4209640, 4209647, 4209667, 4209671, 4209685, 4209694, 4209698, 4209734, 4209764, 4209771, 4209774, 4209775, 4209776, 4209777, 4209778, 4209779, 4209780, 4209781, 4209782, 4209783, 4209784, 4209785, 4209786, 4209787, 4209788, 4209789, 4209790, 4209791, 4209792, 4209793, 4209794, 4209795, 4209796, 4209797, 4209798, 4209799, 4209800, 4209801, 4209802, 4209803, 4209804, 4209805, 4209806, 4209807, 4209808, 4209809, 4209810, 4209811, 4209812, 4209813, 4209814, 4209815, 4209816, 4209817, 4209818, 4209819, 4209820, 4209821, 4209822, 4209823, 4209834, 4209849, 4209859, 4209860, 4209863, 4209927, 4209928, 4209929, 4209930, 4209931, 4209932, 4209933, 4209934, 4209935, 4209936, 4209937, 4209938, 4209939, 4209940, 4209941, 4209942, 4209943, 4209945, 4209946, 4209947, 4209948, 4209949, 4209950, 4209951, 4209952, 4209954, 4209955, 4209956, 4209957, 4209959, 4209960, 4209961, 4209962, 4209963, 4209964, 4209965, 4209966, 4209967, 4209968, 4209969, 4209985, 4209986, 4209987, 4209988 Updated information submitted: Serial numbers removed from 12/23/20 letter: 4200112, 4200117, 4200202, 4201412, 4201490, 4201661; reason: These devices are still in scope and exhibit the issue. However, it was found that none of these devices were distributed to a customer. Therefore, they were removed from the attachment in the letter and are within Stryker control. Serial Numbers added to the recall letter: 4209292, 4209646, 4209841, 4202629, 4203963; reason: These devices were originally thought to be in Stryker control, but were shipped to customers. Serial Numbers added to the recall letter: 4209917, 4209922, 4209924; reason: These devices were shipped to Stryker OUS distributor and were thought to be distributed to customers. (Firm later confirmed that the units are under Stryker control.)
- Product
- Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product…
- Recalling firm
- Mako Surgical Corporation
- Quantity
- 367 units
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK
- Recall date
- November 25, 2020
- Classification
- Class II
- Status
- Completed
- Agency
- FDA (device)