FDAClass IJanuary 19, 2022

Vaporizer Sevoflurane Maquet Filling recall: Impurity Out of Spec

Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia System 6887700…

Why was Vaporizer Sevoflurane Maquet Filling recalled?

A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: 07325710008430 Serial Nos. 15580 15890 15924 16284 16637 16837 16970 15582 15891 15925 16285 16638 16840 16971 15721 15894 16268 16345 16639 16842 15722 15914 16269 16352 16640 16843 15725 15917 16270 16354 16690 16844 15726 15919 16271 16632 16828 16961 15728 15920 16272 16633 16832 16964 15729 15921 16273 16634 16833 16965 15731 15922 16276 16635 16834 16966 15732 15923 16283 16636 16836 16968

Product
Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia System 6887700…
Recalling firm
Getinge Usa Sales Inc
Quantity
50 US; 898 OUS
Distribution
Worldwide distribution - US Nationwide distribution in the states of VA and MO. The countries of Austria, Bahrain, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Estonia, Iraq, Israel, Italy, Japan, Kenya, Malaysia, Mexico, Morocco, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Recall date
January 19, 2022
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls