FDAClass IINovember 20, 2024

IMMULITE 2000 PSA assay recall

IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)

Why was IMMULITE 2000 PSA assay recalled?

The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU).

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

SMN 10706281: UDI - 00630414982212, Lot Number D164; SMN 10706282: UDI - 00630414982229, Lot Number D166; SMN 10380896 (OUS): UDI - 00630414961958; SMN 10380996 (OUS): UDI - 00630414961965; OUS Lot Numbers: 438, 439, 440

Product
IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
7,332 units
Distribution
Worldwide - US Nationwide distribution.
Recall date
November 20, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls