FDAClass IIDecember 3, 2014

Abbott m2000sp recall: Software Issue

Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing.

Why was Abbott m2000sp recalled?

Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

List Number: 09K14; Models: E series, G series

Product
Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing.
Recalling firm
Abbott Molecular
Quantity
9 CD ROMs (US); 1,273 CD ROMs (Foreign)
Distribution
Worldwide distribution. US nationwide (CA, CT, FL, NC, PA, SC, and DC), Algeria , Australia , Austria , Belgium , Bosnia and Herzegovina , Cameroon , Canada , Chad , Colombia , Costa Rica , Croatia , Denmark , Egypt , Ethiopia , Finland , France , Germany , Greece , Guinea , Honduras , Hong Kong , India , Indonesia , Ireland , Israel , Italy , Japan , Kenya , Korea , Lithuania , Mali , Mexico , Netherlands , New Zealand , Nigeria , Norway , Pakistan , Paraguay , Poland , Portugal , Romania , Russia , Saudi Arabia , Senegal , Slovenia , South Africa , Spain , Sweden , Switzerland , Taiwan , Tanzania , Thailand , Turkey , Uganda , Ukraine , United Arab Emirates, United Kingdom , Vietnam
Recall date
December 3, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)