ProxiDiagnost N90 recall
ProxiDiagnost N90; Model Number: 706110;
Why was ProxiDiagnost N90 recalled?
Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Number: 706110; UDI-DI: 00884838103221; Serial Number: 10001185, 10001205, 10001220, 10001224, 10001225, 10001227, 10001228, 10001229, 10001230, 10001231, 10001234, 10001235, 10001236, 10001237, 10001239, 10001245, 10001247, 10001248, 10001249, 10001250, 10001251, 10001253, 10001256, 10001257, 10001258, 10001261, 10001262, 10001264, 10001265, 10001266, 10001269, 10001270, 10001271, 10001272, 10001273, 10001274, 10001275, 10001276, 10001277, 10001278, 10001282;
- Product
- ProxiDiagnost N90; Model Number: 706110;
- Recalling firm
- Philips Medical Systems DMC GmbH
- Quantity
- 41 units
- Distribution
- Worldwide Distribution: US (Nationwide) Distribution and OUS (International) countries of: Australia, Austria, Bahrain, Belgium, Bosnia-Herz., Canada, Chile, Croatia, Denmark, Estonia, Finland, France, Germany, Ghana, Gibraltar, Guadeloupe, Indonesia, Iraq, Israel, Italy, Jordan, Kuwait, Kyrgyzstan, Latvia, Lithuania, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Samoa, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, T¿rkiye, Turkmenistan, United Kingdom, Utd.Arab.Emir.,
- Recall date
- November 19, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)