FDAClass IINovember 19, 2025

Brand Name: Sonesta Product Name: 6210 recall

Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Description: Sonesta 6210 - Procedure table…

Why was Brand Name: Sonesta Product Name: 6210 recalled?

Video fluoroscopy table may contain inadequate weld on backrest bracket resulting in backrest dropping.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Code: Primary DI Number: 07350006950018 Serial numbers of potentially affected units: 12-70791 12-70792 12-70793 12-70794 12-70795 12-70796 12-70797 12-70798 12-70799 12-70801

Product
Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Description: Sonesta 6210 - Procedure table…
Recalling firm
Sonesta Medical AB
Quantity
10 tables
Distribution
US Distribution to NH only.
Recall date
November 19, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls