ARTIS Q.zen biplane with Material 10848355 - a recall
ARTIS Q.zen biplane with Material 10848355 - a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for…
Why was ARTIS Q.zen biplane with Material 10848355 - a recalled?
Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: Serial 123202 123211 123208 123209 123206 123207 123219 123205 123201 123055 123056 123216 123200 123212
- Product
- ARTIS Q.zen biplane with Material 10848355 - a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for…
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 15
- Distribution
- US Nationwide Distribution
- Recall date
- November 21, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)