FDAClass IIDecember 24, 2014

Discovery IGS 740 Indicated for use for patients from recall

Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for…

Why was Discovery IGS 740 Indicated for use for patients from recalled?

Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Mfg Lot or Serial # System ID 00000646778BU1 190020RX26

Product
Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for…
Recalling firm
GE Healthcare, LLC
Quantity
1 - OUS
Distribution
Worldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated