Anesthesia circuit kit recall
Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230
Why was Anesthesia circuit kit recalled?
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230 UDI-DI 40884389688589 (case), 10884389688588 (ea) Lots 25EBV570 25FBT843
- Product
- Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230
- Recalling firm
- Medline Industries, LP
- Quantity
- 96,406 total units
- Distribution
- International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
- Recall date
- November 19, 2025
- Classification
- Class I
- Status
- Ongoing