FDAClass INovember 19, 2025

Dynj909119g Robotic Gastric LAP Sleeve-lf Dynj909127f recall

DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF

Why was Dynj909119g Robotic Gastric LAP Sleeve-lf Dynj909127f recalled?

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

CYSTO DYNJ909127F UDI-DI 40198459055424 (case), 10198459055423 (ea) Lots 25FBP819 25GBM605; GENERAL LAPAROSCOPY-LF DYNJ909129D UDI-DI 40198459055264 (case), 10198459055263 (ea) Lots 25FBV537; ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909119G UDI-DI 40198459130527 (case), 10198459130526 (ea) Lots 25FBP681

Product
DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF
Recalling firm
Medline Industries, LP
Quantity
96,406 total units
Distribution
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Recall date
November 19, 2025
Classification
Class I
Status
Ongoing
Agency
FDA (device)