Blood Monitoring Unit recall
Blood Monitoring Unit (BMU 40), model no. 701040852 - Product Usage: is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar…
Why was Blood Monitoring Unit recalled?
The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial numbers 90002022 90002134 90002163 90002187 90002276 90002297 90002023 90002135 90002164 90002188 90002277 90002298 90002024 90002136 90002168 90002191 90002279 90002302 90002029 90002137 90002170 90002194 90002280 90002306 90002047 90002140 90002172 90002195 90002281 90002312 90002112 90002143 90002181 90002196 90002287 90002313 90002123 90002144 90002183 90002210 90002289 90002124 90002160 90002184 90002212 90002293 90002133 90002161 90002185 90002273 90002296
- Product
- Blood Monitoring Unit (BMU 40), model no. 701040852 - Product Usage: is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar…
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Quantity
- 51 units
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AZ, CA, IA, MD, NC, PA, TX, WA, WI and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, Ecuador, Finland, France, Germany, Hong Kong, India, Iran, Iraq, Ireland, Italy, Jordan, Malaysia, Malta, Netherlands, Norway, Pakistan, Peru, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela.
- Recall date
- November 18, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)