FDAClass IIDecember 5, 2012

Cooper Surgical Advincula Arch recall

Cooper Surgical Advincula Arch , Reusable Uterine Manipulator Handle Ref: UMH750 The intended use of the RUMI Arch is in an operative endoscopy…

Why was Cooper Surgical Advincula Arch recalled?

Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Arch handles with a serial number less than 1204001, or between 1111001 0 and 1111 0069

Product
Cooper Surgical Advincula Arch , Reusable Uterine Manipulator Handle Ref: UMH750 The intended use of the RUMI Arch is in an operative endoscopy…
Recalling firm
Cooper Surgical, Inc.
Quantity
1215 total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Australia, Belgium, Canada, Czech Republic, Germany, Italy, Spain, France, United Kingdom, Japan, New Zealand, South Korea, Saudi Arabia, Sweden, Singapore, Slovenia, and Taiwan.
Recall date
December 5, 2012
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls