220/240V AC recall
220/240V AC, Advanced Perfusion System Platform (APS) The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the…
Why was 220/240V AC recalled?
Terumo Cardiovascular System (TCVS) has received reports of a situation where users experienced a total loss of functionality for some System 1 units. The reports indicate that the units went blank and shut down with no sign of power and battery backup did not initiate. The result is all pumps stop, with no safety system functionality, and the battery would not be activated. The user would be
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog number: 801764 and all serial numbers
- Product
- 220/240V AC, Advanced Perfusion System Platform (APS) The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the…
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Quantity
- 1647 units total
- Distribution
- Worldwide Distribution--USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV.
- Recall date
- December 5, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)