FDAClass IIDecember 4, 2013

13mm addEASE(TM) Binary Connector with 17 Ga. Needle recall

13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial…

Why was 13mm addEASE(TM) Binary Connector with 17 Ga. Needle recalled?

Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Material BC1300, Lot number 61286479.

Product
13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial…
Recalling firm
B. Braun Medical, Inc.
Quantity
4,000 units
Distribution
Nationwide Distribution
Recall date
December 4, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls